An electromagnetic water meter has no moving part in the water. It measures the voltage that moving water induces across a magnetic field — microvolts, in a pipe that is electrically connected to everything around it. Almost every field failure of this type traces back to that second half of the sentence, not to the first.
This is a reading of one integrated battery-powered unit: the DN40–300 electromagnetic water meter, ISO 4064 construction with factory inspection to ISO/IEC 17025, 4G Cat.1 with RS485 for local access. Every figure comes from that document. Where a statement is my inference or a reading of the table rather than a printed sentence, it is marked as such.
Three things on this sheet are easy to misread, and all three change what you order. The range figure R is a configuration rather than a fixed property of each size. Q1 — the minimum flow — is not printed at all; it is derived. And the size series has a gap in the middle that decides a whole band of duties.
What the tube measures, and why the signal is measured in microvolts
Faraday's law of induction, applied to a pipe: a conductor moving through a magnetic field develops an electromotive force proportional to its velocity, the flux density and the distance between the electrodes. In a meter, the conductor is the water, the field comes from a pair of coils, and the electrodes sit across the tube. The result is a signal proportional to mean axial velocity, and therefore to volume flow, with nothing in the water to wear or jam.
The proportionality constant is small and the voltage is smaller. Working back from a typical water velocity of a metre or two per second and a tube of a few hundred millimetres, the electrode signal sits in the tens to hundreds of microvolts. That single fact explains most of the installation rules on this datasheet: a signal that small cannot tolerate a competing voltage, and it cannot be referenced to nothing.
It also explains the feature list. There is no strainer requirement, no straight-pipe conditioner and no bearing to wear, which is why the manual can claim performance "not affected by water impurities" in the sense that sand passing through does not abrade a rotor — there is not one. The tolerance is for conductive water, which is a different constraint and the subject of the next section.
The 5 µS/cm threshold is the real application limit
The specification states a medium conductivity of at least 5 µS/cm. That number is the boundary of the technology, not a quality grade, and it is worth knowing what sits on each side of it.
Municipal drinking water, borehole water and most raw surface water sit far above it — typically by two to three orders of magnitude. Almost any water a distribution network carries will operate an electromagnetic meter without difficulty.
Below the line you find the fluids that look like water and are not: deionised water, reverse-osmosis permeate, condensate, demineralised loop water, and some process streams after ion exchange. In those, there are too few charge carriers for the induced voltage to rise above the noise floor, and the reading drifts or falls to zero. A meter installed on a line that is later switched to softened or demineralised supply is a classic failure that no amount of recalibration fixes.
If your fluid is anywhere near the threshold, measure the conductivity before specifying rather than after commissioning. The manual gives the limit and no margin figure; a fluid at 6 µS/cm is inside the stated range but has very little headroom against a drifting zero.
The flow table: Q1 is not printed, it is derived
Read the columns carefully, because two of them are formulas rather than numbers. Q1 is given as Q3/R. Q2 is given as 1.6/Q1, which is only dimensionally sensible read as 1.6 × Q1 — the transition ratio the standard fixes, printed here with the operator inverted. The R column, meanwhile, is a single merged cell holding seven values: 630, 500, 400, 315, 250, 200 and 160.
| DN | Q4 overload m³/h | Q3 permanent m³/h | Q2 | Q1 | Resolution | Max reading |
|---|---|---|---|---|---|---|
| 40 | 31.25 | 25 | 1.6 × Q1 | Q3 / R | 0.0001 | 999999.9999 after carry-over 99999999.99 |
| 50 | 50 | 40 | ||||
| 80 | 78.75 | 63 | ||||
| 100 | 125 | 100 | ||||
| 150 | 312.5 | 250 | ||||
| 200 | 500 | 400 | ||||
| 250 | 787.5 | 630 | ||||
| 300 | 1250 | 1000 |
Two arithmetic patterns hold in every row, and they are the same two as on any compliant Class 2 meter: Q4/Q3 is 1.25 throughout, and Q2/Q1 is 1.6 throughout. Neither was chosen per size. What is genuinely size-specific here is Q3 alone — and, unlike a mechanical meter, what R you take.
The register figure deserves a note as well. The table prints 999999.9999 and, underneath, "after carrying over", 99999999.99. The reading that makes sense of the pair is that the display runs six integer digits and four decimals up to roll-over, then widens to eight integer digits with two decimals. If that is right, the resolution available after roll-over is coarser — 0.01 m³ instead of 0.0001. It is a detail that matters on a long-life asset, so confirm it against a meter before relying on the last two decimals.