Not sure this is the right bore? Work it out from your peak and minimum flow — the sizing calculator needs two numbers.
Give us the peak flow you must measure and the smallest flow you must still see, and this works out the bore, the Q3, and which range ratio puts your minimum inside the accuracy band. It uses the relationships printed in the meter documentation, and the published flow tables of this range.
Every figure below comes from the published specification tables of this range. Where a bore has no printed Q3, it is Q4 ÷ 1.25, which is the relationship the documentation states.
Four flows define any meter, and three of them are fixed once you choose Q3 and the ratio. This is the arithmetic the calculator runs:
| Q3 — rated flow | The flow the meter is designed to run at continuously |
| Q4 — overload flow | Q4 = 1.25 × Q3. The meter has to survive it briefly |
| R — range ratio | R = Q3 ÷ Q1. Chosen when the meter is ordered |
| Q1 — minimum flow | Q1 = Q3 ÷ R. The bottom of the accuracy band |
| Q2 — transitional flow | Q2 = 1.6 × Q1. Between Q1 and Q2 the tolerance is wider |
Class 2 accuracy means ±5% between Q1 and Q2, and ±2% between Q2 and Q4. That is why the ratio matters more than the bore: at a fixed Q3, going from R250 to R500 halves Q1 and moves your leakage threshold into the ±2% band.
If the line is a solar thermal loop rather than a mains supply — glycol, high temperature, a pump curve instead of a pressure main — the four typical cases are worked through in our sizing guide for solar thermal flow meters.
| Bore | Q3 m³/h | Q4 m³/h | Ratios offered | Basis |
|---|---|---|---|---|
| DN15 | 2.50 | 3.125 | R80–R500 · R400–R1000 by family | Q4 printed; Q3 = Q4 ÷ 1.25 |
| DN20 | 4.00 | 5.000 | R80–R500 · R400–R1000 by family | Q4 printed; Q3 = Q4 ÷ 1.25 |
| DN25 | 6.30 | 7.875 | R80–R500 · R400–R1000 by family | Q4 printed; Q3 = Q4 ÷ 1.25 |
| DN32 | 10.0 | 12.50 | R160–R630 | Q4 printed; Q3 = Q4 ÷ 1.25 |
| DN40 | 16.0 | 20.00 | R160–R630 | Q4 printed; Q3 = Q4 ÷ 1.25 |
| DN50 | 25.0 | 31.25 | R160–R630 · R250–R500 flanged · R400 standard | Q3 printed (DN50 and LXCY-50 pages) |
| DN65 | 40.0 | 50.00 | R250 / R400 / R500 | Q3 printed; R400 standard |
| DN80 | 63.0 | 78.75 | R250 / R400 / R500 | Q4 printed in the DN50–DN150 family table |
| DN100 | 100 | 125.0 | R250 / R400 / R500 | Q4 printed in the family table |
| DN125 | 160 | 200.0 | R250 / R400 / R500 | Q4 printed in the family table |
| DN150 | 250 | 312.5 | R250 / R400 / R500 | Q4 printed in the family table |
These are the two questions engineers ask immediately after sizing, and they are the two we can only answer to the extent they are published:
New to the flow ratings? Q3, Q4 and the DN you actually need explains what each number decides — and why oversizing a meter costs resolution at night — before you enter anything here.
Give us the peak flow, the smallest flow you must see and the line size. We come back with the bore, the ratio, the datasheet for that model and a price — no form to fill in first.